prEN ISO 24966
Determination of flash point - Modified continuously closed cup flash point (MCCCFP) method (ISO/DIS 24966:2025)
Draft Public enquiryProject stage codes
1. Prioritetinių darbo temų pažymėjimas ir ekspertų skyrimas
2. Projektas
3. Viešoji apklausa
From 2025-05-23
till 2025-07-22
4. Closure of enquiry
5. Submission to Formal Vote
Organisation
CEN Europos standartizacijos komitetasTechnical Committee
TK 40 Petroleuml, petroleum products, lubricants and alternative fuelsForeignTC'S
CEN/TC 19Number of comments
0Comment start date
2025-05-23Scope
This International Standard specifies a test method for the determination of the flash point of chemicals, lube oils, aviation turbine fuel, diesel fuel, diesel/biodiesel blends and other liquids by a continuously closed cup tester utilizing a specimen size of 2 ml, cup size of 7 ml, with a heating rate of 2.5 °C per minute. This flash point test method is a dynamic method and depends on definite rates of temperature increase. It is one of the many flash point test methods available and every flash point test method, including this one, is an empirical method. It utilises an electric arc as the ignitor and detects the flash point by pressure measurement.This test method is suitable for testing samples with a flash point from 22,5 °C to 235,5 °C. Flash point determinations below 22,5 °C and above 235,5 °C may be performed, but the precision has not been determined.