prEN 16589-2
Laboratory local exhaust devices - Articulated extraction arms - Part 2: Commissioning and on-site testing
Draft Public enquiryProject stage codes
1. Decision on WI Proposal
2. Projektas
3. Viešoji apklausa
From 2025-05-02
till 2025-07-01
4. Closure of enquiry
5. Submission to Formal Vote
Organisation
CEN Europos standartizacijos komitetasICS
71.040.10 - Chemical laboratories. Laboratory equipmentForeignTC'S
CEN/TC 332Number of comments
0Comment start date
2025-05-02Scope
This document applies to test methods designed to be used at the place of installation of the AEA, usually a laboratory and for various laboratory applications that require local extraction. They are used for commissioning after installation, for maintenance and for qualification purposes. For certain customer requirements additional or modified test methods can be necessary. This document includes product functional performance referring to product standard detailed in EN 16589-1. Occupational health and safety assessments methods are not included in this document. This document does not consider performance requirements for the extract air system associated with the AEA installation and therefore extract system performance is not part of the scope. This document does not confirm or establish a capture zone of an AEA capture device only the functional extract capacity and mechanical functions of the AEA.